The BabyGuard W-1140 open reanimation system offers three control modes: pre-warm mode control, manual mode control and baby mode control by micro-computer. The set temperature and baby temperature can be displayed separately and the horizontal angle of the warmer module and inclination of the bassinet are adjustable.

Request Information
Request Price/Quote

Features
• Three control modes: pre-warm mode control, manual mode control and baby mode control by micro-computer
• Set temperature and baby temperature can be displayed separately
• Failure alarm indication
• Horizontal angle of warmer module and inclination of the bassinet is adjustable.
• The panels around bassinet can be turned outward
• X-ray cassette under the bassinet
• Timer to count phototherapy time
• APGAR timer
Specification
• Mode of control: Pre-warm mode, manual mode, baby mode are controlled by micro-computer
• Baby mode temperature control range: 34.5°-37.5°C
• Temperature sensor display range: 5-65°C
• Accuracy of skin temperature sensor: ±0.2°C
• Temperature uniformity of mattress: ≤2°
• Angle of warmer module: 0°, 30°, 60° for two ways
• Inclination of bassinet: ±10°
• Average value of general bilirubin: radiation on the mattress effective range: ≥0.58 mW/cm²
• Bilirubin uniformity of radiating on the mattress effective range: >0.4
• Wave length of blue light: 400-550 nm
• Operating range: 18°C – 30°C
• Ambient air velocity rate: