Transport Infant Incubator | Hospimedica Expo
Summary
|
Products
Avante Medical Surgical (formerly DRE)
Transport Infant Incubator  
Model: Avante NeoGlider
Quick Inquiry
Description
The Avante NeoGlider transport infant incubator features multiple temperature control modes, each controlled by micro-computer, and uses AC power and DC power alternatively, or DC 12V or DC 24V power in an ambulance. The set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately while a failure alarm indication ensures the infant’s safety.

Request Information
Request Price/Quote

Features
• Use AC power and DC power alternatively, and DC 12V or DC 24V power in an ambulance
• Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately
• Can be equipped with >37°C temperature set function
• Failure alarm indication
• Double wall hood with side door for pulling out the bassinet
• Natural air flow humidity
• Adjustable height
Specification
• AC power supply: AC110V - 240V/50-60Hz
• DC power supply: DC12V/10A or DC24V/6A
• Input power: 400VA
• Air-temperature control range: 25° - 37°C / Override mode: 37° - 38°C
• Baby temperature control range: 34° - 37°C / Override mode: 37° - 37.5°C
• Skin temperature sensor accuracy: ≤0.3° C
• Temperature variability: ≤1.0° C
• Temperature uniformity: ≤1.5° C (mattress is on the horizontal position)
• Noise inside hood: ≤55dB(A)
• Environment noise ≤45dB(A)
• Failure alarm: Over-temperature alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure alarm
• Operating at temperature equilibrium: 90 min (Set temp: 36°C | Ambient temperature: 15°C)
• Size Main body: 109 cm × 65.5 cm × 76 cm
• Size Cabinet: 138 cm × 61 cm × 40 cm
• Gross Weight Main unit: 80 kg
• Gross Weight Cabinet: 53 kg

Request Information