The MEDRAD Avanta fluid management injection system for X-ray angiography was specifically designed with the contrast administration needs of interventional cardiology in mind. It allows controlled delivery of radiopaque contrast media and common flushing solutions for interventional imaging purposes. The injector facilitates contrast delivery in fixed or variable flow mode and saline flushing at a fixed rate. It utilizes dedicated disposables that include multi-patient components and single-patient components.
Request Information
Request Price/Quote
Specification
• Flow Rates:
Contrast: Fixed Flow: User defined from 1–45 mL/sec in increments of 0.1 mL/sec;
Variable Flow: User controlled variable range from 1 to 10 mL/sec in 0.1 increments
Saline: Fixed Rate: 1.0 mL/sec
• Volume: User controlled variable range 1–150 mL in increments of 1 mL
• Pressure Limits: User defined from 300 PSI to 1,200 PSI in increments of 1 PSI
• Fill Rate: Manual or automatic refill 2 mL/sec
• Injection Delay: 0.1 sec to 0–99.9 sec
• Rise Time: Programmable at 0.1 sec increments from 0.1 to 9.9 sec
• Memory: Up to 40 customizable protocols store
• Display Screen: 26 cm (~10.5") Color Touch Screen
• Certegra™ ISI Capable: Provides for synchronization of the injection and X-ray exposure