Proton Therapy Dosimetry Head Phantom
Model:Proton Therapy Dosimetry Head Model 731-HN
Description
The CIRS Proton Therapy Dosimetry Head is an anthropomorphic head phantom designed for commissioning and treatment planning system (TPS) verification with any conformal or IMRT Proton Therapy system. The phantom is constructed of CIRS tissue-equivalent materials, which mimic reference tissues within 1.5% for protons and within 1% for photons from 50 KeV to 15 MeV.
Features
• Detailed internal anatomy including bone and air
• Three film locations in sagittal direction
• Tissue Equivalent for protons and photons
• Dental filling and spine prothesis